4-Channel Source Measurement Unit
The NI PXIe-4141 is a high-channel-count, high-speed, precision source measure unit (SMU) with four identical SMU channels per PXI Express slot.
Each SMU offers 4-quadrant operation to source or sink ±10 V at ±100 mA. With integrated remote (4-wire) sensing and current measurement sensitivity of 10 pA, the NI PXIe-4141 is ideal for precision measurements including accurate characterisation of idle currents.
Specifications:
General | |
Product Family | Power Supplies |
Form Factor | PXI Platform |
PXI Bus Type | PXI Express |
Part Number | 781743-01 |
Operating System/Target | Real-Time , Windows |
LabVIEW RT Support | Yes |
Triggering | SoftwareTrigger , ExternalTrigger , InternalTrigger |
External Clocking | No |
I/O Connector | 25-pin D-Sub |
Product Name | NI PXIe-4141 |
Analog Output | |
Channels | 4 |
Resolution | 18 bits |
Timing | Hardware |
Update Rate | 100 kS/s |
Current Drive Single | 100 mA |
Current Drive All | 400 mA |
Output Impedance | -100 kOhm , 100 kOhm |
Max Voltage | -10 V , 10 V |
Maximum Voltage Range Accuracy | 600 µV |
Maximum Voltage Range Sensitivity | 10 µV |
Minimum Voltage Range | -10 V , 10 V |
Minimum Voltage Range Accuracy | 600 µV |
Min Voltage Range Sensitivity | 10 µV |
Max Current Range | -100 mA , 100 mA |
Maximum Current Range Accuracy | 15 µA |
Maximum Current Range Sensitivity | 100 nA |
Minimum Current Range | -10 µA , 10 µA |
Minimum Current Range Accuracy | 1.5 nA |
Min Current Range Sensitivity | 10 pA |
Output Quadrants | Quadrant III , Quadrant II , Quadrant IV , Quadrant I |
Max Quadrants Per Channel | 4 |
To find out more about the NI PXIe, kindly log on to http://sine.ni.com/ds/app/doc/p/id/ds-359/lang/en
To contact National Instruments Australia about 4-Channel Source Measurement Unit use Get a quote.
The NI PXIe-4141 is a high-channel-count, high-speed, precision source measure unit (SMU) with four identical SMU channels per PXI Express slot.
Each SMU offers 4-quadrant operation to source or sink ±10 V at ±100 mA. With integrated remote (4-wire) sensing and current measurement sensitivity of 10 pA, the NI PXIe-4141 is ideal for precision measurements including accurate characterisation of idle currents.
Specifications:
General | |
Product Family | Power Supplies |
Form Factor | PXI Platform |
PXI Bus Type | PXI Express |
Part Number | 781743-01 |
Operating System/Target | Real-Time , Windows |
LabVIEW RT Support | Yes |
Triggering | SoftwareTrigger , ExternalTrigger , InternalTrigger |
External Clocking | No |
I/O Connector | 25-pin D-Sub |
Product Name | NI PXIe-4141 |
Analog Output | |
Channels | 4 |
Resolution | 18 bits |
Timing | Hardware |
Update Rate | 100 kS/s |
Current Drive Single | 100 mA |
Current Drive All | 400 mA |
Output Impedance | -100 kOhm , 100 kOhm |
Max Voltage | -10 V , 10 V |
Maximum Voltage Range Accuracy | 600 µV |
Maximum Voltage Range Sensitivity | 10 µV |
Minimum Voltage Range | -10 V , 10 V |
Minimum Voltage Range Accuracy | 600 µV |
Min Voltage Range Sensitivity | 10 µV |
Max Current Range | -100 mA , 100 mA |
Maximum Current Range Accuracy | 15 µA |
Maximum Current Range Sensitivity | 100 nA |
Minimum Current Range | -10 µA , 10 µA |
Minimum Current Range Accuracy | 1.5 nA |
Min Current Range Sensitivity | 10 pA |
Output Quadrants | Quadrant III , Quadrant II , Quadrant IV , Quadrant I |
Max Quadrants Per Channel | 4 |